| 48 T | $412.00 |
| 96 T | $588.00 |
| 96 T*5 | $2499.00 |
| Concentration (pg/mL) | OD | Corrected OD |
|---|---|---|
| 2000.00 | 2.131 | 2.049 |
| 1000.00 | 1.629 | 1.547 |
| 500.00 | 1.227 | 1.145 |
| 250.00 | 0.928 | 0.846 |
| 125.00 | 0.457 | 0.375 |
| 62.50 | 0.357 | 0.275 |
| 31.25 | 0.179 | 0.097 |
| 0.00 | 0.082 | 0.000 |

Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
| Matrix | Recovery range | Average |
|---|---|---|
| serum(n=5) | 87-99% | 93% |
| EDTA plasma(n=5) | 83-99% | 95% |
| Heparin plasma(n=5) | 80-96% | 88% |
| Matrix | 1:2 | 1:4 | 1:8 | 1:16 |
|---|---|---|---|---|
| serum(n=5) | 86-97% | 93-102% | 87-98% | 83-97% |
| EDTA plasma(n=5) | 82-93% | 95-104% | 82-96% | 85-98% |
| Heparin plasma(n=5) | 85-94% | 87-96% | 92-101% | 82-98% |