| 48 T | $412.00 |
| 96 T | $588.00 |
| 96 T*5 | $2499.00 |
| Concentration (pg/mL) | OD | Corrected OD |
|---|---|---|
| 1000.00 | 1.842 | 1.734 |
| 500.00 | 1.542 | 1.434 |
| 250.00 | 1.109 | 1.001 |
| 125.00 | 0.868 | 0.760 |
| 62.50 | 0.591 | 0.483 |
| 31.25 | 0.339 | 0.231 |
| 15.63 | 0.207 | 0.099 |
| 0.00 | 0.108 | 0.000 |

Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
| Matrix | Recovery range | Average |
|---|---|---|
| serum(n=5) | 86-99% | 92% |
| EDTA plasma(n=5) | 83-97% | 90% |
| Heparin plasma(n=5) | 89-103% | 96% |
| Matrix | 1:2 | 1:4 | 1:8 | 1:16 |
|---|---|---|---|---|
| serum(n=5) | 89-99% | 83-96% | 98-104% | 88-101% |
| EDTA plasma(n=5) | 79-91% | 96-105% | 80-92% | 91-103% |
| Heparin plasma(n=5) | 92-104% | 87-96% | 89-101% | 82-90% |