48 T | $412.00 |
96 T | $588.00 |
Concentration (pg/mL) | OD | Corrected OD |
---|---|---|
1000.00 | 2.126 | 2.040 |
500.00 | 1.630 | 1.544 |
250.00 | 1.035 | 0.949 |
125.00 | 0.847 | 0.761 |
62.50 | 0.539 | 0.453 |
31.25 | 0.301 | 0.215 |
15.63 | 0.234 | 0.148 |
0.00 | 0.086 | 0.000 |
Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
Matrix | Recovery range | Average |
---|---|---|
serum(n=5) | 83-97% | 91% |
EDTA plasma(n=5) | 82-94% | 88% |
Heparin plasma(n=5) | 94-105% | 98% |
Matrix | 1:2 | 1:4 | 1:8 | 1:16 |
---|---|---|---|---|
serum(n=5) | 91-99% | 88-97% | 88-104% | 97-108% |
EDTA plasma(n=5) | 93-102% | 86-92% | 96-105% | 89-97% |
Heparin plasma(n=5) | 85-100% | 84-98% | 97-101% | 88-96% |