| 48 T | $412.00 |
| 96 T | $588.00 |
| 96 T*5 | $2499.00 |
| Concentration (pg/mL) | OD | Corrected OD |
|---|---|---|
| 1000.00 | 1.941 | 1.852 |
| 500.00 | 1.565 | 1.476 |
| 250.00 | 1.128 | 1.039 |
| 125.00 | 0.855 | 0.766 |
| 62.50 | 0.513 | 0.424 |
| 31.25 | 0.318 | 0.229 |
| 15.63 | 0.135 | 0.046 |
| 0.00 | 0.089 | 0.000 |

Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
| Matrix | Recovery range | Average |
|---|---|---|
| serum(n=5) | 90-105% | 97% |
| EDTA plasma(n=5) | 86-99% | 92% |
| Heparin plasma(n=5) | 79-93% | 86% |
| Matrix | 1:2 | 1:4 | 1:8 | 1:16 |
|---|---|---|---|---|
| serum(n=5) | 93-102% | 86-92% | 96-105% | 89-97% |
| EDTA plasma(n=5) | 90-99% | 83-97% | 81-103% | 89-101% |
| Heparin plasma(n=5) | 79-91% | 95-104% | 93-102% | 79-94% |