| 48 T | $412.00 |
| 96 T | $588.00 |
| 96 T*5 | $2499.00 |
| Concentration (pg/mL) | OD | Corrected OD |
|---|---|---|
| 1000.00 | 2.195 | 2.114 |
| 500.00 | 1.527 | 1.446 |
| 250.00 | 1.106 | 1.025 |
| 125.00 | 0.958 | 0.877 |
| 62.50 | 0.545 | 0.464 |
| 31.25 | 0.357 | 0.276 |
| 15.63 | 0.221 | 0.140 |
| 0.00 | 0.081 | 0.000 |

Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
| Matrix | Recovery range | Average |
|---|---|---|
| serum(n=5) | 90-107% | 98% |
| EDTA plasma(n=5) | 87-99% | 93% |
| Heparin plasma(n=5) | 82-98% | 90% |
| Matrix | 1:2 | 1:4 | 1:8 | 1:16 |
|---|---|---|---|---|
| serum(n=5) | 78-95% | 96-103% | 83-101% | 82-97% |
| EDTA plasma(n=5) | 86-93% | 93-101% | 86-97% | 90-98% |
| Heparin plasma(n=5) | 91-99% | 83-89% | 88-94% | 87-93% |