| 48 T | $320.00 |
| 96 T | $458.00 |
| 96 T*5 | $1947.00 |
| Concentration (pg/mL) | OD | Corrected OD |
|---|---|---|
| 1000.00 | 2.270 | 2.202 |
| 500.00 | 1.609 | 1.541 |
| 250.00 | 1.136 | 1.068 |
| 125.00 | 0.857 | 0.789 |
| 62.50 | 0.543 | 0.475 |
| 31.25 | 0.363 | 0.295 |
| 15.63 | 0.225 | 0.157 |
| 0.00 | 0.068 | 0.000 |

Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
| Matrix | Recovery range | Average |
|---|---|---|
| serum(n=5) | 95-107% | 101% |
| EDTA plasma(n=5) | 85-95% | 90% |
| Heparin plasma(n=5) | 79-93% | 86% |
| Matrix | 1:2 | 1:4 | 1:8 | 1:16 |
|---|---|---|---|---|
| serum(n=5) | 83-103% | 82-93% | 86-95% | 81-96% |
| EDTA plasma(n=5) | 86-96% | 85-97% | 89-101% | 85-97% |
| Heparin plasma(n=5) | 80-101% | 87-96% | 96-105% | 95-103% |