| 48 T | $412.00 |
| 96 T | $588.00 |
| 96 T*5 | $2499.00 |
| Concentration (pg/mL) | OD | Corrected OD |
|---|---|---|
| 1000.00 | 2.164 | 2.075 |
| 500.00 | 1.782 | 1.693 |
| 250.00 | 1.183 | 1.094 |
| 125.00 | 0.937 | 0.848 |
| 62.50 | 0.532 | 0.443 |
| 31.25 | 0.369 | 0.280 |
| 15.63 | 0.194 | 0.105 |
| 0.00 | 0.089 | 0.000 |

Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
| Matrix | Recovery range | Average |
|---|---|---|
| serum(n=5) | 83-97% | 90% |
| EDTA plasma(n=5) | 81-95% | 88% |
| Heparin plasma(n=5) | 85-95% | 90% |
| Matrix | 1:2 | 1:4 | 1:8 | 1:16 |
|---|---|---|---|---|
| serum(n=5) | 78-97% | 86-97% | 81-103% | 88-97% |
| EDTA plasma(n=5) | 96-104% | 90-102% | 87-98% | 95-103% |
| Heparin plasma(n=5) | 88-97% | 96-102% | 81-93% | 89-97% |