| 48 T | $320.00 |
| 96 T | $458.00 |
| 96 T*5 | $1947.00 |
| Concentration (ng/mL) | OD | Corrected OD |
|---|---|---|
| 10.00 | 2.080 | 1.981 |
| 5.00 | 1.779 | 1.680 |
| 2.50 | 1.237 | 1.138 |
| 1.25 | 0.806 | 0.707 |
| 0.63 | 0.567 | 0.468 |
| 0.31 | 0.360 | 0.261 |
| 0.16 | 0.202 | 0.103 |
| 0.00 | 0.099 | 0.000 |

Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
| Matrix | Recovery range | Average |
|---|---|---|
| serum(n=5) | 92-119% | 106% |
| EDTA plasma(n=5) | 83-128% | 106% |
| Heparin plasma(n=5) | 91-104% | 98% |
| Matrix | 1:2 | 1:4 | 1:8 | 1:16 |
|---|---|---|---|---|
| serum(n=5) | 95-111% | 82-112% | 93-99% | 95-105% |
| EDTA plasma(n=5) | 93-112% | 91-105% | 92-118% | 82-88% |
| Heparin plasma(n=5) | 84-114% | 95-101% | 88-117% | 87-98% |