| 48 T | $320.00 |
| 96 T | $458.00 |
| 96 T*5 | $1947.00 |
| Concentration (ng/mL) | OD | Corrected OD |
|---|---|---|
| 10.00 | 2.167 | 2.086 |
| 5.00 | 1.592 | 1.511 |
| 2.50 | 1.257 | 1.176 |
| 1.25 | 0.854 | 0.773 |
| 0.63 | 0.511 | 0.430 |
| 0.31 | 0.289 | 0.208 |
| 0.16 | 0.219 | 0.138 |
| 0.00 | 0.081 | 0.000 |

Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
| Matrix | Recovery range | Average |
|---|---|---|
| serum(n=5) | 92-117% | 105% |
| EDTA plasma(n=5) | 88-129% | 109% |
| Heparin plasma(n=5) | 89-109% | 99% |
| Matrix | 1:2 | 1:4 | 1:8 | 1:16 |
|---|---|---|---|---|
| serum(n=5) | 86-115% | 87-105% | 81-102% | 85-118% |
| EDTA plasma(n=5) | 84-111% | 85-110% | 88-99% | 87-103% |
| Heparin plasma(n=5) | 82-93% | 87-115% | 82-106% | 93-107% |