| 48 T | $320.00 |
| 96 T | $458.00 |
| 96 T*5 | $1947.00 |
| Concentration (pg/mL) | OD | Corrected OD |
|---|---|---|
| 2000.00 | 1.915 | 1.819 |
| 1000.00 | 1.548 | 1.452 |
| 500.00 | 1.176 | 1.080 |
| 250.00 | 0.971 | 0.875 |
| 125.00 | 0.582 | 0.486 |
| 62.50 | 0.370 | 0.274 |
| 31.25 | 0.211 | 0.115 |
| 0.00 | 0.096 | 0.000 |

Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
| Matrix | Recovery range | Average |
|---|---|---|
| serum(n=5) | 81-89% | 85% |
| EDTA plasma(n=5) | 92-109% | 101% |
| Heparin plasma(n=5) | 84-106% | 95% |
| Matrix | 1:2 | 1:4 | 1:8 | 1:16 |
|---|---|---|---|---|
| serum(n=5) | 84-101% | 82-98% | 92-112% | 84-113% |
| EDTA plasma(n=5) | 89-113% | 86-98% | 83-99% | 87-108% |
| Heparin plasma(n=5) | 87-115% | 81-114% | 92-112% | 81-98% |