| 48 T | $320.00 |
| 96 T | $458.00 |
| 96 T*5 | $1947.00 |
| Concentration (ng/mL) | OD | Corrected OD |
|---|---|---|
| 10.00 | 1.993 | 1.905 |
| 5.00 | 1.613 | 1.525 |
| 2.50 | 1.175 | 1.087 |
| 1.25 | 0.825 | 0.737 |
| 0.63 | 0.530 | 0.442 |
| 0.31 | 0.255 | 0.167 |
| 0.16 | 0.154 | 0.066 |
| 0.00 | 0.088 | 0.000 |

Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
| Matrix | Recovery range | Average |
|---|---|---|
| serum(n=5) | 91-112% | 102% |
| EDTA plasma(n=5) | 93-122% | 108% |
| Heparin plasma(n=5) | 90-109% | 100% |
| Matrix | 1:2 | 1:4 | 1:8 | 1:16 |
|---|---|---|---|---|
| serum(n=5) | 91-104% | 92-110% | 91-111% | 80-102% |
| EDTA plasma(n=5) | 80-93% | 93-120% | 93-104% | 81-106% |
| Heparin plasma(n=5) | 86-99% | 84-97% | 94-102% | 94-102% |