Products

Human LAB7-1(B-Lymphocyte Activation Antigen B7-1) Microsample ELISA Kit
One-click to copy product information
ELK5326MS
Size: Price:
48 T $320.00
96 T $458.00
96 T*5 $1947.00

Overview

Product name: Human LAB7-1(B-Lymphocyte Activation Antigen B7-1) Microsample ELISA Kit
Reactivity: Human
Alternative Names: CD80; CD28LG; CD28LG1; BB1; B7-1 Antigen; T-lymphocyte activation antigen CD80; CTLA-4 counter-receptor B7.1
Assay Type: Sandwich
Sensitivity: 32 pg/mL
Standard: 5000 pg/mL
Detection Range: 78.13-5000 pg/mL
Sample Type: Serum, plasma, tissue homogenates and other biological fluids
Assay Length: 3.5h
Research Area: CD & Adhesion molecule;Tumor immunity;Infection immunity;Immune molecule;Autoimmunity;
Test principle: The test principle applied in this kit is Sandwich enzyme immunoassay. The microtiter plate provided in this kit has been pre-coated with an antibody specific to Human LAB7-1. Standards or samples are added to the appropriate microtiter plate wells then with a biotin-conjugated antibody specific to Human LAB7-1. Next, Avidin conjugated to Horseradish Peroxidase (HRP) is added to each microplate well and incubated. After TMB substrate solution is added, only those wells that contain Human LAB7-1, biotin-conjugated antibody and enzyme-conjugated Avidin will exhibit a change in color. The enzyme-substrate reaction is terminated by the addition of sulphuric acid solution and the color change is measured spectrophotometrically at a wavelength of 450nm ± 10nm. The concentration of Human LAB7-1 in the samples is then determined by comparing the OD of the samples to the standard curve.

Standard curve

Concentration (pg/mL) OD Corrected OD
5000.00 2.249 2.165
2500.00 1.800 1.716
1250.00 1.247 1.163
625.00 0.856 0.772
312.50 0.472 0.388
156.25 0.301 0.217
78.13 0.229 0.145
0.00 0.084 0.000

Precision

Intra-assay Precision (Precision within an assay):CV%<8%

Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.

Inter-assay Precision (Precision between assays):CV%<10%

Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.

Recovery

Matrices listed below were spiked with certain level of recombinant LAB7-1 and the recovery rates were calculated by comparing the measured value to the expected amount of LAB7-1 in samples.
Matrix Recovery range Average
serum(n=5) 94-120% 107%
EDTA plasma(n=5) 90-106% 98%
Heparin plasma(n=5) 86-119% 103%

Linearity

The linearity of the kit was assayed by testing samples spiked with appropriate concentration of LAB7-1 and their serial dilutions. The results were demonstrated by the percentage of calculated concentration to the expected.
Matrix 1:2 1:4 1:8 1:16
serum(n=5) 89-115% 89-110% 94-113% 94-119%
EDTA plasma(n=5) 95-107% 90-114% 88-104% 91-99%
Heparin plasma(n=5) 91-113% 86-105% 90-99% 80-102%

Product Components