48 T | $320.00 |
96 T | $458.00 |
Concentration (pg/mL) | OD | Corrected OD |
---|---|---|
10000.00 | 0.212 | |
5000.00 | 0.386 | |
2500.00 | 0.557 | |
1250.00 | 0.857 | |
625.00 | 1.156 | |
312.50 | 1.511 | |
156.25 | 1.715 | |
0.00 | 2.211 |
Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
Matrix | Recovery range | Average |
---|---|---|
serum(n=5) | 88-102% | 95% |
EDTA plasma(n=5) | 83-98% | 90% |
Heparin plasma(n=5) | 81-95% | 86% |
Matrix | 1:2 | 1:4 | 1:8 | 1:16 |
---|---|---|---|---|
serum(n=5) | 88-103% | 83-94% | 84-97% | 89-107% |
EDTA plasma(n=5) | 95-105% | 82-95% | 81-97% | 93-105% |
Heparin plasma(n=5) | 86-95% | 85-95% | 89-99% | 85-95% |