Size: | Price: |
---|---|
48 T | $320.00 |
96 T | $458.00 |
Concentration (ng/mL) | OD | Corrected OD |
---|---|---|
1000.00 | 1.932 | 1.849 |
500.00 | 1.674 | 1.591 |
250.00 | 1.174 | 1.091 |
125.00 | 0.828 | 0.745 |
62.50 | 0.556 | 0.473 |
31.25 | 0.351 | 0.268 |
15.63 | 0.176 | 0.093 |
0.00 | 0.083 | 0.000 |
Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
Matrix | Recovery range | Average |
---|---|---|
serum(n=5) | 82-95% | 88% |
EDTA plasma(n=5) | 84-97% | 90% |
Heparin plasma(n=5) | 79-93% | 86% |
Matrix | 1:2 | 1:4 | 1:8 | 1:16 |
---|---|---|---|---|
serum(n=5) | 85-97% | 87-96% | 95-103% | 88-95% |
EDTA plasma(n=5) | 86-93% | 93-101% | 86-97% | 90-101% |
Heparin plasma(n=5) | 78-92% | 90-101% | 87-98% | 85-92% |