48 T | $320.00 |
96 T | $458.00 |
Concentration (pg/mL) | OD | Corrected OD |
---|---|---|
1000.00 | 0.199 | |
500.00 | 0.385 | |
250.00 | 0.567 | |
125.00 | 0.895 | |
62.50 | 1.168 | |
31.25 | 1.537 | |
15.63 | 1.624 | |
0.00 | 2.028 |
Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
Matrix | Recovery range | Average |
---|---|---|
serum(n=5) | 85-98% | 91% |
EDTA plasma(n=5) | 78-90% | 84% |
Heparin plasma(n=5) | 95-107% | 98% |
Matrix | 1:2 | 1:4 | 1:8 | 1:16 |
---|---|---|---|---|
serum(n=5) | 85-94% | 87-99% | 89-100% | 85-96% |
EDTA plasma(n=5) | 86-95% | 93-103% | 85-94% | 87-101% |
Heparin plasma(n=5) | 95-102% | 90-96% | 87-1028% | 85-97% |