| 48 T | $320.00 |
| 96 T | $458.00 |
| 96 T*5 | $1947.00 |
| Concentration (pg/mL) | OD | Corrected OD |
|---|---|---|
| 1000.00 | 2.079 | 1.996 |
| 500.00 | 1.638 | 1.555 |
| 250.00 | 1.121 | 1.038 |
| 125.00 | 0.797 | 0.714 |
| 62.50 | 0.572 | 0.489 |
| 31.25 | 0.332 | 0.249 |
| 15.63 | 0.165 | 0.082 |
| 0.00 | 0.083 | 0.000 |

Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
| Matrix | Recovery range | Average |
|---|---|---|
| serum(n=5) | 84-98% | 91% |
| EDTA plasma(n=5) | 79-97% | 88% |
| Heparin plasma(n=5) | 80-95% | 87% |
| Matrix | 1:2 | 1:4 | 1:8 | 1:16 |
|---|---|---|---|---|
| serum(n=5) | 82-94% | 85-96% | 78-90% | 91-99% |
| EDTA plasma(n=5) | 80-92% | 86-102% | 87-103% | 85-101% |
| Heparin plasma(n=5) | 82-93% | 96-105% | 87-98% | 85-97% |