| 48 T | $320.00 |
| 96 T | $458.00 |
| 96 T*5 | $1947.00 |
| Concentration (pg/mL) | OD | Corrected OD |
|---|---|---|
| 2000.00 | 2.051 | 1.963 |
| 1000.00 | 1.628 | 1.540 |
| 500.00 | 1.219 | 1.131 |
| 250.00 | 0.901 | 0.813 |
| 125.00 | 0.527 | 0.439 |
| 62.50 | 0.369 | 0.281 |
| 31.25 | 0.216 | 0.128 |
| 0.00 | 0.088 | 0.000 |

Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
| Matrix | Recovery range | Average |
|---|---|---|
| serum(n=5) | 80-95% | 87% |
| EDTA plasma(n=5) | 83-98% | 90% |
| Heparin plasma(n=5) | 86-99% | 92% |
| Matrix | 1:2 | 1:4 | 1:8 | 1:16 |
|---|---|---|---|---|
| serum(n=5) | 79-94% | 87-96% | 85-98% | 95-104% |
| EDTA plasma(n=5) | 85-103% | 82-96% | 87-105% | 90-99% |
| Heparin plasma(n=5) | 88-102% | 86-92% | 81-93% | 79-87% |