| 48 T | $320.00 |
| 96 T | $458.00 |
| 96 T*5 | $1947.00 |
| Concentration (pg/mL) | OD | Corrected OD |
|---|---|---|
| 500.00 | 1.944 | 1.842 |
| 250.00 | 1.609 | 1.507 |
| 125.00 | 1.156 | 1.054 |
| 62.50 | 0.912 | 0.810 |
| 31.25 | 0.512 | 0.410 |
| 15.63 | 0.393 | 0.291 |
| 7.81 | 0.226 | 0.124 |
| 0.00 | 0.102 | 0.000 |

Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
| Matrix | Recovery range | Average |
|---|---|---|
| serum(n=5) | 89-110% | 100% |
| EDTA plasma(n=5) | 92-109% | 101% |
| Heparin plasma(n=5) | 92-119% | 106% |
| Matrix | 1:2 | 1:4 | 1:8 | 1:16 |
|---|---|---|---|---|
| serum(n=5) | 81-95% | 85-106% | 86-103% | 90-99% |
| EDTA plasma(n=5) | 82-99% | 94-106% | 92-99% | 81-100% |
| Heparin plasma(n=5) | 86-104% | 93-108% | 91-116% | 94-114% |