| 48 T | $320.00 |
| 96 T | $458.00 |
| 96 T*5 | $1947.00 |
| Concentration (pg/mL) | OD | Corrected OD |
|---|---|---|
| 500.00 | 1.996 | 1.893 |
| 250.00 | 1.577 | 1.474 |
| 125.00 | 1.144 | 1.041 |
| 62.50 | 0.933 | 0.830 |
| 31.25 | 0.513 | 0.410 |
| 15.63 | 0.385 | 0.282 |
| 7.82 | 0.221 | 0.118 |
| 0.00 | 0.103 | 0.000 |

Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
| Matrix | Recovery range | Average |
|---|---|---|
| serum(n=5) | 90-105% | 97% |
| EDTA plasma(n=5) | 95-107% | 101% |
| Heparin plasma(n=5) | 78-90% | 84% |
| Matrix | 1:2 | 1:4 | 1:8 | 1:16 |
|---|---|---|---|---|
| serum(n=5) | 98-107% | 90-99% | 88-97% | 85-97% |
| EDTA plasma(n=5) | 81-94% | 87-105% | 79-95% | 96-105% |
| Heparin plasma(n=5) | 82-94% | 95-102% | 87-103% | 88-95% |