Size: | Price: |
---|---|
48 T | $320.00 |
96 T | $458.00 |
Concentration (ng/mL) | OD | Corrected OD |
---|---|---|
10.00 | 2.045 | 1.962 |
5.00 | 1.597 | 1.514 |
2.50 | 1.136 | 1.053 |
1.25 | 0.752 | 0.669 |
0.63 | 0.529 | 0.446 |
0.32 | 0.362 | 0.279 |
0.16 | 0.170 | 0.087 |
0.00 | 0.083 | 0.000 |
Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
Matrix | Recovery range | Average |
---|---|---|
serum(n=5) | 80-95% | 87% |
EDTA plasma(n=5) | 85-98% | 91% |
Heparin plasma(n=5) | 96-107% | 102% |
Matrix | 1:2 | 1:4 | 1:8 | 1:16 |
---|---|---|---|---|
serum(n=5) | 95-103% | 86-92% | 89-97% | 91-103% |
EDTA plasma(n=5) | 85-94% | 87-101% | 86-97% | 88-95% |
Heparin plasma(n=5) | 85-92% | 79-96% | 92-101% | 87-98% |