| 48 T | $320.00 |
| 96 T | $458.00 |
| 96 T*5 | $1947.00 |
| Concentration (ng/mL) | OD | Corrected OD |
|---|---|---|
| 10.00 | 2.189 | 2.090 |
| 5.00 | 1.482 | 1.383 |
| 2.50 | 1.194 | 1.095 |
| 1.25 | 0.828 | 0.729 |
| 0.63 | 0.508 | 0.409 |
| 0.32 | 0.388 | 0.289 |
| 0.16 | 0.225 | 0.126 |
| 0.00 | 0.099 | 0.000 |

Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
| Matrix | Recovery range | Average |
|---|---|---|
| serum(n=5) | 80-92% | 86% |
| EDTA plasma(n=5) | 92-106% | 99% |
| Heparin plasma(n=5) | 87-99% | 93% |
| Matrix | 1:2 | 1:4 | 1:8 | 1:16 |
|---|---|---|---|---|
| serum(n=5) | 89-99% | 83-99% | 98-105% | 96-101% |
| EDTA plasma(n=5) | 91-102% | 86-102% | 87-99% | 99-105% |
| Heparin plasma(n=5) | 86-93% | 91-101% | 87-98% | 95-104% |