| 48 T | $412.00 |
| 96 T | $588.00 |
| 96 T*5 | $2499.00 |
| Concentration (pg/mL) | OD | Corrected OD |
|---|---|---|
| 3000.00 | 2.126 | 2.044 |
| 1500.00 | 1.572 | 1.490 |
| 750.00 | 1.216 | 1.134 |
| 375.00 | 0.796 | 0.714 |
| 187.50 | 0.479 | 0.397 |
| 93.75 | 0.403 | 0.321 |
| 46.88 | 0.215 | 0.133 |
| 0.00 | 0.082 | 0.000 |

Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
| Matrix | Recovery range | Average |
|---|---|---|
| serum(n=5) | 85-99% | 92% |
| EDTA plasma(n=5) | 93-107% | 100% |
| Heparin plasma(n=5) | 87-99% | 93% |
| Matrix | 1:2 | 1:4 | 1:8 | 1:16 |
|---|---|---|---|---|
| serum(n=5) | 85-94% | 87-99% | 85-101% | 79-90% |
| EDTA plasma(n=5) | 88-97% | 96-103% | 86-92% | 89-97% |
| Heparin plasma(n=5) | 92-104% | 86-98% | 80-93% | 89-97% |