| 48 T | $412.00 |
| 96 T | $588.00 |
| 96 T*5 | $2499.00 |
| Concentration (ng/mL) | OD | Corrected OD |
|---|---|---|
| 100.00 | 2.185 | 2.096 |
| 50.00 | 1.615 | 1.526 |
| 25.00 | 1.152 | 1.063 |
| 12.50 | 0.963 | 0.874 |
| 6.25 | 0.593 | 0.504 |
| 3.13 | 0.342 | 0.253 |
| 1.57 | 0.215 | 0.126 |
| 0.00 | 0.089 | 0.000 |

Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
| Matrix | Recovery range | Average |
|---|---|---|
| serum(n=5) | 83-92% | 85% |
| EDTA plasma(n=5) | 84-108% | 98% |
| Heparin plasma(n=5) | 87-96% | 92% |
| Matrix | 1:2 | 1:4 | 1:8 | 1:16 |
|---|---|---|---|---|
| serum(n=5) | 95-104% | 82-96% | 92-105% | 83-92% |
| EDTA plasma(n=5) | 91-98% | 83-101% | 87-98% | 95-102% |
| Heparin plasma(n=5) | 80-97% | 81-92% | 87-103% | 82-93% |