| 48 T | $320.00 |
| 96 T | $458.00 |
| 96 T*5 | $1947.00 |
| Concentration (pg/mL) | OD | Corrected OD |
|---|---|---|
| 2000.00 | 2.328 | 2.234 |
| 1000.00 | 1.643 | 1.549 |
| 500.00 | 1.220 | 1.126 |
| 250.00 | 0.670 | 0.576 |
| 125.00 | 0.387 | 0.293 |
| 62.50 | 0.297 | 0.203 |
| 31.25 | 0.229 | 0.135 |
| 0.00 | 0.094 | 0.000 |

Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
| Matrix | Recovery range | Average |
|---|---|---|
| serum(n=5) | 82-88% | 85% |
| EDTA plasma(n=5) | 84-114% | 99% |
| Heparin plasma(n=5) | 89-118% | 104% |
| Matrix | 1:2 | 1:4 | 1:8 | 1:16 |
|---|---|---|---|---|
| serum(n=5) | 93-114% | 91-114% | 85-95% | 80-107% |
| EDTA plasma(n=5) | 93-98% | 94-113% | 93-118% | 85-97% |
| Heparin plasma(n=5) | 95-120% | 83-96% | 86-98% | 85-105% |