48 T | $320.00 |
96 T | $458.00 |
Concentration (pg/mL) | OD | Corrected OD |
---|---|---|
2000.00 | 0.183 | |
1000.00 | 0.375 | |
500.00 | 0.596 | |
250.00 | 0.842 | |
125.00 | 1.139 | |
62.50 | 1.585 | |
31.25 | 1.796 | |
0.00 | 2.256 |
Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
Matrix | Recovery range | Average |
---|---|---|
serum(n=5) | 87-99% | 93% |
EDTA plasma(n=5) | 80-95% | 87% |
Heparin plasma(n=5) | 87-99% | 93% |
Matrix | 1:2 | 1:4 | 1:8 | 1:16 |
---|---|---|---|---|
serum(n=5) | 81-94% | 85-97% | 91-102% | 81-90% |
EDTA plasma(n=5) | 83-97% | 92-103% | 86-97% | 93-101% |
Heparin plasma(n=5) | 83-95% | 87-96% | 79-91% | 95-103% |