Size: | Price: |
---|---|
48 T | $320.00 |
96 T | $458.00 |
Concentration (pg/mL) | OD | Corrected OD |
---|---|---|
1000.00 | 2.156 | 2.057 |
500.00 | 1.755 | 1.656 |
250.00 | 1.135 | 1.036 |
125.00 | 0.752 | 0.653 |
62.50 | 0.567 | 0.468 |
31.25 | 0.313 | 0.214 |
15.63 | 0.196 | 0.097 |
0.00 | 0.099 | 0.000 |
Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
Matrix | Recovery range | Average |
---|---|---|
serum(n=5) | 95-108% | 101% |
EDTA plasma(n=5) | 83-96% | 89% |
Heparin plasma(n=5) | 88-102% | 95% |
Matrix | 1:2 | 1:4 | 1:8 | 1:16 |
---|---|---|---|---|
serum(n=5) | 88-101% | 85-93% | 89-97% | 85-94% |
EDTA plasma(n=5) | 87-103% | 97-105% | 79-90% | 87-96% |
Heparin plasma(n=5) | 85-96% | 87-101% | 88-102% | 87-102% |