Products

High Sensitivity Rat I-PTH(IntactParathormone) ELISA Kit
One-click to copy product information
ELK2563HS
Size: Price:
48 T $320.00
96 T $458.00
96 T*5 $1947.00

Overview

Product name: High Sensitivity Rat I-PTH(IntactParathormone) ELISA Kit
Reactivity: Rat
Alternative Names: iPTH; Intact Parathyroid Hormone; Parathormone; Parathyrin; PTH; Parathyroid Hormone
Assay Type: Sandwich
Sensitivity: 1.5 pg/mL
Standard: 200 pg/mL
Detection Range: 3.13-200 pg/mL
Sample Type: Serum, plasma
Assay Length: 3.5h
Research Area: Metabolic pathway;Endocrinology;Hormone metabolism;
Test principle: The test principle applied in this kit is Sandwich enzyme immunoassay. The microtiter plate provided in this kit has been pre-coated with an antibody specific to Rat I-PTH. Standards or samples are added to the appropriate microtiter plate wells then with a biotin-conjugated antibody specific to Rat I-PTH. Next, Avidin conjugated to Horseradish Peroxidase (HRP) is added to each microplate well and incubated. After TMB substrate solution is added, only those wells that contain Rat I-PTH, biotin-conjugated antibody and enzyme-conjugated Avidin will exhibit a change in color. The enzyme-substrate reaction is terminated by the addition of sulphuric acid solution and the color change is measured spectrophotometrically at a wavelength of 450nm ± 10nm. The concentration of Rat I-PTH in the samples is then determined by comparing the OD of the samples to the standard curve.

Standard curve

Concentration (pg/mL) OD Corrected OD
200.00 2.379 2.291
100.00 1.563 1.475
50.00 1.213 1.125
25.00 0.861 0.773
12.50 0.508 0.420
6.25 0.335 0.247
3.13 0.189 0.101
0.00 0.088 0.000

Precision

Intra-assay Precision (Precision within an assay):CV%<8%

Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.

Inter-assay Precision (Precision between assays):CV%<10%

Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.

Recovery

Matrices listed below were spiked with certain level of recombinant I-PTH and the recovery rates were calculated by comparing the measured value to the expected amount of I-PTH in samples.
Matrix Recovery range Average
serum(n=5) 83-97% 90%
EDTA plasma(n=5) 95-103% 99%
Heparin plasma(n=5) 91-100% 96%

Linearity

The linearity of the kit was assayed by testing samples spiked with appropriate concentration of I-PTH and their serial dilutions. The results were demonstrated by the percentage of calculated concentration to the expected.
Matrix 1:2 1:4 1:8 1:16
serum(n=5) 87-97% 94-106% 82-90% 96-106%
EDTA plasma(n=5) 89-102% 83-95% 91-101% 91-101%
Heparin plasma(n=5) 92-103% 95-101% 88-99% 79-86%