| 48 T | $320.00 |
| 96 T | $458.00 |
| 96 T*5 | $1947.00 |
| Concentration (pg/mL) | OD | Corrected OD |
|---|---|---|
| 50.00 | 3.027 | 2.941 |
| 25.00 | 2.145 | 2.059 |
| 12.50 | 1.481 | 1.395 |
| 6.25 | 0.943 | 0.857 |
| 3.13 | 0.508 | 0.422 |
| 1.57 | 0.251 | 0.165 |
| 0.79 | 0.146 | 0.060 |
| 0.00 | 0.086 | 0.000 |

Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
| Matrix | Recovery range | Average |
|---|---|---|
| serum(n=5) | 95-101% | 98% |
| EDTA plasma(n=5) | 81-93% | 87% |
| Heparin plasma(n=5) | 83-97% | 90% |
| Matrix | 1:2 | 1:4 | 1:8 | 1:16 |
|---|---|---|---|---|
| serum(n=5) | 79-86% | 81-88% | 78-86% | 81-93% |
| EDTA plasma(n=5) | 88-100% | 88-99% | 95-108% | 95-108% |
| Heparin plasma(n=5) | 88-94% | 93-106% | 80-94% | 92-100% |