| 48 T | $320.00 |
| 96 T | $458.00 |
| 96 T*5 | $1947.00 |
| Concentration (pg/mL) | OD | Corrected OD |
|---|---|---|
| 1000.00 | 2.153 | 2.049 |
| 500.00 | 1.628 | 1.524 |
| 250.00 | 1.128 | 1.024 |
| 125.00 | 0.901 | 0.797 |
| 62.50 | 0.523 | 0.419 |
| 31.25 | 0.312 | 0.208 |
| 15.63 | 0.228 | 0.124 |
| 0.00 | 0.104 | 0.000 |

Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
| Matrix | Recovery range | Average |
|---|---|---|
| serum(n=5) | 87-95% | 91% |
| EDTA plasma(n=5) | 78-92% | 85% |
| Heparin plasma(n=5) | 92-106% | 99% |
| Matrix | 1:2 | 1:4 | 1:8 | 1:16 |
|---|---|---|---|---|
| serum(n=5) | 86-93% | 91-105% | 87-98% | 99-102% |
| EDTA plasma(n=5) | 89-99% | 83-101% | 93-104% | 96-101% |
| Heparin plasma(n=5) | 85-107% | 84-98% | 95-103% | 86-98% |