| 48 T | $320.00 |
| 96 T | $458.00 |
| 96 T*5 | $1947.00 |
| Concentration (pg/mL) | OD | Corrected OD |
|---|---|---|
| 1000.00 | 2.266 | 2.169 |
| 500.00 | 1.563 | 1.466 |
| 250.00 | 1.073 | 0.976 |
| 125.00 | 0.839 | 0.742 |
| 62.50 | 0.453 | 0.356 |
| 31.25 | 0.317 | 0.220 |
| 15.63 | 0.232 | 0.135 |
| 0.00 | 0.097 | 0.000 |

Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
| Matrix | Recovery range | Average |
|---|---|---|
| serum(n=5) | 80-95% | 87% |
| EDTA plasma(n=5) | 85-98% | 91% |
| Heparin plasma(n=5) | 96-107% | 102% |
| Matrix | 1:2 | 1:4 | 1:8 | 1:16 |
|---|---|---|---|---|
| serum(n=5) | 92-107% | 83-99% | 80-96% | 91-101% |
| EDTA plasma(n=5) | 85-94% | 87-99% | 85-101% | 79-90% |
| Heparin plasma(n=5) | 86-95% | 93-103% | 87-98% | 85-96% |