Products

Human MPIF1(Myeloid Progenitor Inhibitory Factor 1) ELISA Kit
ELK2475
Size: Price:
48 T $320.00
96 T $458.00

Overview

Product name: Human MPIF1(Myeloid Progenitor Inhibitory Factor 1) ELISA Kit
Reactivity: Human
Alternative Names: CCL23; SCYA23; Ckb-8; MIP-3; CKb8; Chemokine(C-C-Motif)ligand 23; Small Inducible Cytokine Subfamily A(Cys-Cys)Member 23; Macrophage inflammatory protein 3
Assay Type: Sandwich
Sensitivity: 0.47 pg/mL
Standard: 1000 pg/mL
Detection Range: 15.63-1000 pg/mL
Sample Type: Serum, plasma and other biological fluids.
Assay Length: 3.5h
Research Area: Cytokine;Tumor immunity;Infection immunity;
Uniprot ID: P55773
Test principle: The test principle applied in this kit is Sandwich enzyme immunoassay. The microtiter plate provided in this kit has been pre-coated with an antibody specific to Human MPIF1. Standards or samples are added to the appropriate microtiter plate wells then with a biotin-conjugated antibody specific to Human MPIF1. Next, Avidin conjugated to Horseradish Peroxidase (HRP) is added to each microplate well and incubated. After TMB substrate solution is added, only those wells that contain Human MPIF1, biotin-conjugated antibody and enzyme-conjugated Avidin will exhibit a change in color. The enzyme-substrate reaction is terminated by the addition of sulphuric acid solution and the color change is measured spectrophotometrically at a wavelength of 450nm ± 10nm. The concentration of Human MPIF1 in the samples is then determined by comparing the OD of the samples to the standard curve.

Standard curve

Concentration (pg/mL) OD Corrected OD
1000.00 2.157 2.064
500.00 1.472 1.379
250.00 1.075 0.982
125.00 0.775 0.682
62.50 0.474 0.381
31.25 0.265 0.172
15.63 0.235 0.142
0.00 0.093 0.000

Precision

Intra-assay Precision (Precision within an assay):CV%<8%

Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.

Inter-assay Precision (Precision between assays):CV%<10%

Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.

Recovery

Matrices listed below were spiked with certain level of recombinant MPIF1 and the recovery rates were calculated by comparing the measured value to the expected amount of MPIF1 in samples.
Matrix Recovery range Average
serum(n=5) 81-95% 88%
EDTA plasma(n=5) 82-97% 89%
Heparin plasma(n=5) 95-107% 101%

Linearity

The linearity of the kit was assayed by testing samples spiked with appropriate concentration of MPIF1 and their serial dilutions. The results were demonstrated by the percentage of calculated concentration to the expected.
Matrix 1:2 1:4 1:8 1:16
serum(n=5) 86-96% 93-104% 89-97% 85-94%
EDTA plasma(n=5) 87-96% 85-97% 85-94% 87-96%
Heparin plasma(n=5) 92-101% 82-98% 89-97% 85-93%