Size: | Price: |
---|---|
48 T | $320.00 |
96 T | $458.00 |
Concentration (pg/mL) | OD | Corrected OD |
---|---|---|
4000.00 | 2.194 | 2.108 |
2000.00 | 1.586 | 1.500 |
1000.00 | 1.223 | 1.137 |
500.00 | 0.907 | 0.821 |
250.00 | 0.518 | 0.432 |
125.00 | 0.387 | 0.301 |
62.50 | 0.252 | 0.166 |
0.00 | 0.086 | 0.000 |
Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
Matrix | Recovery range | Average |
---|---|---|
serum(n=5) | 95-107% | 101% |
EDTA plasma(n=5) | 96-104% | 100% |
Heparin plasma(n=5) | 98-104% | 102% |
Matrix | 1:2 | 1:4 | 1:8 | 1:16 |
---|---|---|---|---|
serum(n=5) | 89-99% | 88-104% | 84-98% | 95-105% |
EDTA plasma(n=5) | 86-97% | 83-96% | 91-103% | 96-105% |
Heparin plasma(n=5) | 95-102% | 98-106% | 87-98% | 91-102% |