| 48 T | $320.00 |
| 96 T | $458.00 |
| 96 T*5 | $1947.00 |
| Concentration (pg/mL) | OD | Corrected OD |
|---|---|---|
| 1000.00 | 2.255 | 2.166 |
| 500.00 | 1.768 | 1.679 |
| 250.00 | 1.087 | 0.998 |
| 125.00 | 0.815 | 0.726 |
| 62.50 | 0.549 | 0.460 |
| 31.25 | 0.317 | 0.228 |
| 15.63 | 0.210 | 0.121 |
| 0.00 | 0.089 | 0.000 |

Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
| Matrix | Recovery range | Average |
|---|---|---|
| serum(n=5) | 83-115% | 99% |
| EDTA plasma(n=5) | 84-121% | 103% |
| Heparin plasma(n=5) | 86-117% | 102% |
| Matrix | 1:2 | 1:4 | 1:8 | 1:16 |
|---|---|---|---|---|
| serum(n=5) | 94-115% | 92-118% | 81-99% | 84-98% |
| EDTA plasma(n=5) | 87-107% | 83-105% | 90-95% | 90-97% |
| Heparin plasma(n=5) | 84-109% | 87-99% | 87-116% | 83-116% |