| 48 T | $320.00 |
| 96 T | $458.00 |
| 96 T*5 | $1947.00 |
| Concentration (pg/mL) | OD | Corrected OD |
|---|---|---|
| 1000.00 | 2.265 | 2.198 |
| 500.00 | 1.609 | 1.542 |
| 250.00 | 1.157 | 1.090 |
| 125.00 | 0.853 | 0.786 |
| 62.50 | 0.500 | 0.433 |
| 31.25 | 0.359 | 0.292 |
| 15.63 | 0.230 | 0.163 |
| 0.00 | 0.067 | 0.000 |

Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
| Matrix | Recovery range | Average |
|---|---|---|
| serum(n=5) | 91-114% | 103% |
| EDTA plasma(n=5) | 92-106% | 99% |
| Heparin plasma(n=5) | 95-105% | 100% |
| Matrix | 1:2 | 1:4 | 1:8 | 1:16 |
|---|---|---|---|---|
| serum(n=5) | 88-118% | 92-99% | 95-114% | 84-95% |
| EDTA plasma(n=5) | 93-117% | 89-110% | 85-96% | 86-112% |
| Heparin plasma(n=5) | 87-105% | 84-114% | 85-104% | 81-97% |