| 48 T | $320.00 | 
| 96 T | $458.00 | 
| 96 T*5 | $1947.00 | 
| Concentration (pg/mL) | OD | Corrected OD | 
|---|---|---|
| 1500.00 | 0.221 | |
| 750.00 | 0.387 | |
| 375.00 | 0.692 | |
| 187.50 | 0.933 | |
| 93.75 | 1.253 | |
| 46.88 | 1.587 | |
| 23.44 | 1.871 | |
| 0.00 | 2.265 | 

Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
| Matrix | Recovery range | Average | 
|---|---|---|
| serum(n=5) | 78-93% | 85% | 
| EDTA plasma(n=5) | 87-99% | 93% | 
| Heparin plasma(n=5) | 90-103% | 96% | 
| Matrix | 1:2 | 1:4 | 1:8 | 1:16 | 
|---|---|---|---|---|
| serum(n=5) | 85-94% | 97-103% | 82-90% | 92-101% | 
| EDTA plasma(n=5) | 95-102% | 97-105% | 92-101% | 85-92% | 
| Heparin plasma(n=5) | 92-104% | 82-96% | 95-102% | 87-98% |