| 48 T | $320.00 |
| 96 T | $458.00 |
| 96 T*5 | $1947.00 |
| Concentration (ng/mL) | OD | Corrected OD |
|---|---|---|
| 10.00 | 2.266 | 2.178 |
| 5.00 | 1.563 | 1.475 |
| 2.50 | 1.073 | 0.985 |
| 1.25 | 0.882 | 0.794 |
| 0.63 | 0.548 | 0.460 |
| 0.32 | 0.262 | 0.174 |
| 0.16 | 0.216 | 0.128 |
| 0.00 | 0.088 | 0.000 |

Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
| Matrix | Recovery range | Average |
|---|---|---|
| serum(n=5) | 87-99% | 93% |
| EDTA plasma(n=5) | 80-95% | 87% |
| Heparin plasma(n=5) | 81-95% | 88% |
| Matrix | 1:2 | 1:4 | 1:8 | 1:16 |
|---|---|---|---|---|
| serum(n=5) | 88-99% | 81-93% | 89-97% | 85-94% |
| EDTA plasma(n=5) | 92-106% | 92-101% | 80-95% | 87-96% |
| Heparin plasma(n=5) | 85-96% | 87-101% | 89-102% | 85-98% |