| 48 T | $320.00 | 
| 96 T | $458.00 | 
| 96 T*5 | $1947.00 | 
| Concentration (pg/mL) | OD | Corrected OD | 
|---|---|---|
| 2000.00 | 2.379 | 2.294 | 
| 1000.00 | 1.657 | 1.572 | 
| 500.00 | 1.226 | 1.141 | 
| 250.00 | 0.834 | 0.749 | 
| 125.00 | 0.541 | 0.456 | 
| 62.50 | 0.336 | 0.251 | 
| 31.25 | 0.175 | 0.090 | 
| 0.00 | 0.085 | 0.000 | 

Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
| Matrix | Recovery range | Average | 
|---|---|---|
| serum(n=5) | 97-105% | 101% | 
| EDTA plasma(n=5) | 83-99% | 95% | 
| Heparin plasma(n=5) | 85-97% | 91% | 
| Matrix | 1:2 | 1:4 | 1:8 | 1:16 | 
|---|---|---|---|---|
| serum(n=5) | 88-104% | 84-98% | 95-105% | 88-102% | 
| EDTA plasma(n=5) | 88-95% | 98-107% | 90-99% | 88-97% | 
| Heparin plasma(n=5) | 92-106% | 93-105% | 85-99% | 90-99% |