| 48 T | $320.00 |
| 96 T | $458.00 |
| 96 T*5 | $1947.00 |
| Concentration (pg/mL) | OD | Corrected OD |
|---|---|---|
| 1000.00 | 1.964 | 1.869 |
| 500.00 | 1.612 | 1.517 |
| 250.00 | 1.176 | 1.081 |
| 125.00 | 0.943 | 0.848 |
| 62.50 | 0.621 | 0.526 |
| 31.25 | 0.354 | 0.259 |
| 15.63 | 0.232 | 0.137 |
| 0.00 | 0.095 | 0.000 |

Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
| Matrix | Recovery range | Average |
|---|---|---|
| serum(n=5) | 88-103% | 95% |
| EDTA plasma(n=5) | 77-95% | 86% |
| Heparin plasma(n=5) | 92-106% | 99% |
| Matrix | 1:2 | 1:4 | 1:8 | 1:16 |
|---|---|---|---|---|
| serum(n=5) | 79-95% | 96-105% | 85-96% | 90-99% |
| EDTA plasma(n=5) | 83-97% | 81-103% | 89-101% | 81-98% |
| Heparin plasma(n=5) | 85-101% | 93-104% | 87-103% | 91-102% |