Size: | Price: |
---|---|
48 T | $265.00 |
96 T | $376.00 |
Concentration (ng/mL) | OD | Corrected OD |
---|---|---|
10.00 | 2.402 | 2.326 |
5.00 | 1.863 | 1.787 |
2.50 | 1.148 | 1.072 |
1.25 | 0.878 | 0.802 |
0.63 | 0.579 | 0.503 |
0.32 | 0.369 | 0.293 |
0.16 | 0.214 | 0.138 |
0.00 | 0.076 | 0.000 |
Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
Matrix | Recovery range | Average |
---|---|---|
serum(n=5) | 78-95% | 86% |
EDTA plasma(n=5) | 80-96% | 88% |
Heparin plasma(n=5) | 80-92% | 86% |
Matrix | 1:2 | 1:4 | 1:8 | 1:16 |
---|---|---|---|---|
serum(n=5) | 89-99% | 83-103% | 98-105% | 96-101% |
EDTA plasma(n=5) | 85-105% | 84-98% | 95-102% | 88-102% |
Heparin plasma(n=5) | 86-93% | 91-105% | 87-98% | 99-105% |