| 48 T | $265.00 |
| 96 T | $376.00 |
| 96 T*5 | $1598.00 |
| Concentration (ng/mL) | OD | Corrected OD |
|---|---|---|
| 10.00 | 0.112 | 0.000 |
| 5.00 | 0.258 | 0.000 |
| 2.50 | 0.426 | 0.000 |
| 1.25 | 0.734 | 0.000 |
| 0.63 | 1.333 | 0.000 |
| 0.32 | 1.900 | 0.000 |
| 0.16 | 2.539 | 0.000 |
| 0.00 | 3.178 | 0.000 |

Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
| Matrix | Recovery range | Average |
|---|---|---|
| serum(n=5) | 90-97% | 94% |
| EDTA plasma(n=5) | 84-91% | 88% |
| Heparin plasma(n=5) | 87-99% | 93% |
| Matrix | 1:2 | 1:4 | 1:8 | 1:16 |
|---|---|---|---|---|
| serum(n=5) | 90-104% | 92-99% | 84-90% | 92-100% |
| EDTA plasma(n=5) | 89-95% | 83-93% | 88-98% | 88-98% |
| Heparin plasma(n=5) | 86-97% | 80-87% | 89-98% | 95-106% |