| 48 T | $265.00 |
| 96 T | $376.00 |
| 96 T*5 | $1598.00 |
| Concentration (ng/mL) | OD | Corrected OD |
|---|---|---|
| 10.00 | 0.142 | 0.000 |
| 5.00 | 0.254 | 0.000 |
| 2.50 | 0.347 | 0.000 |
| 1.25 | 0.687 | 0.000 |
| 0.63 | 1.150 | 0.000 |
| 0.32 | 1.617 | 0.000 |
| 0.16 | 2.150 | 0.000 |
| 0.00 | 2.886 | 0.000 |

Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
| Matrix | Recovery range | Average |
|---|---|---|
| serum(n=5) | 80-88% | 84% |
| EDTA plasma(n=5) | 78-90% | 84% |
| Heparin plasma(n=5) | 92-105% | 99% |
| Matrix | 1:2 | 1:4 | 1:8 | 1:16 |
|---|---|---|---|---|
| serum(n=5) | 83-94% | 93-100% | 94-103% | 89-98% |
| EDTA plasma(n=5) | 86-96% | 78-87% | 93-107% | 93-107% |
| Heparin plasma(n=5) | 88-97% | 91-99% | 86-99% | 81-87% |