| 48 T | $320.00 |
| 96 T | $458.00 |
| 96 T*5 | $1947.00 |
| Concentration (ng/mL) | OD | Corrected OD |
|---|---|---|
| 10.00 | 2.956 | 2.877 |
| 5.00 | 2.197 | 2.118 |
| 2.50 | 1.498 | 1.419 |
| 1.25 | 1.036 | 0.957 |
| 0.63 | 0.656 | 0.577 |
| 0.32 | 0.366 | 0.287 |
| 0.16 | 0.170 | 0.091 |
| 0.00 | 0.079 | 0.000 |

Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
| Matrix | Recovery range | Average |
|---|---|---|
| serum(n=5) | 81-93% | 87% |
| EDTA plasma(n=5) | 81-95% | 88% |
| Heparin plasma(n=5) | 93-100% | 97% |
| Matrix | 1:2 | 1:4 | 1:8 | 1:16 |
|---|---|---|---|---|
| serum(n=5) | 82-92% | 89-102% | 96-103% | 85-95% |
| EDTA plasma(n=5) | 94-104% | 91-105% | 89-100% | 89-100% |
| Heparin plasma(n=5) | 90-98% | 95-104% | 96-102% | 94-107% |