| 48 T | $265.00 |
| 96 T | $376.00 |
| 96 T*5 | $1598.00 |
| Concentration (pg/mL) | OD | Corrected OD |
|---|---|---|
| 2000.00 | 2.485 | 2.424 |
| 1000.00 | 1.249 | 1.188 |
| 500.00 | 0.656 | 0.595 |
| 250.00 | 0.326 | 0.265 |
| 125.00 | 0.192 | 0.131 |
| 62.50 | 0.111 | 0.050 |
| 31.25 | 0.073 | 0.012 |
| 0.00 | 0.061 | 0.000 |

Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
| Matrix | Recovery range | Average |
|---|---|---|
| serum(n=5) | 86-95% | 91% |
| EDTA plasma(n=5) | 78-90% | 84% |
| Heparin plasma(n=5) | 95-106% | 101% |
| Matrix | 1:2 | 1:4 | 1:8 | 1:16 |
|---|---|---|---|---|
| serum(n=5) | 82-95% | 94-100% | 92-101% | 95-102% |
| EDTA plasma(n=5) | 84-96% | 95-104% | 83-94% | 83-94% |
| Heparin plasma(n=5) | 79-93% | 83-92% | 94-105% | 81-90% |