| 48 T | $265.00 |
| 96 T | $376.00 |
| 96 T*5 | $1598.00 |
| Concentration (pg/mL) | OD | Corrected OD |
|---|---|---|
| 500.00 | 3.374 | 3.311 |
| 250.00 | 1.708 | 1.645 |
| 125.00 | 1.094 | 1.031 |
| 62.50 | 0.588 | 0.525 |
| 31.25 | 0.275 | 0.212 |
| 15.63 | 0.140 | 0.077 |
| 7.82 | 0.073 | 0.010 |
| 0.00 | 0.063 | 0.000 |

Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
| Matrix | Recovery range | Average |
|---|---|---|
| serum(n=5) | 90-98% | 94% |
| EDTA plasma(n=5) | 90-96% | 93% |
| Heparin plasma(n=5) | 85-95% | 90% |
| Matrix | 1:2 | 1:4 | 1:8 | 1:16 |
|---|---|---|---|---|
| serum(n=5) | 91-100% | 79-89% | 96-105% | 87-100% |
| EDTA plasma(n=5) | 88-99% | 96-106% | 91-104% | 91-104% |
| Heparin plasma(n=5) | 77-89% | 77-87% | 77-83% | 79-90% |