| 48 T | $265.00 |
| 96 T | $376.00 |
| 96 T*5 | $1598.00 |
| Concentration (ng/mL) | OD | Corrected OD |
|---|---|---|
| 10.00 | 2.817 | 2.738 |
| 5.00 | 1.802 | 1.723 |
| 2.50 | 0.990 | 0.911 |
| 1.25 | 0.625 | 0.546 |
| 0.63 | 0.426 | 0.347 |
| 0.32 | 0.288 | 0.209 |
| 0.16 | 0.155 | 0.076 |
| 0.00 | 0.079 | 0.000 |

Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
| Matrix | Recovery range | Average |
|---|---|---|
| serum(n=5) | 77-90% | 84% |
| EDTA plasma(n=5) | 78-84% | 81% |
| Heparin plasma(n=5) | 95-108% | 102% |
| Matrix | 1:2 | 1:4 | 1:8 | 1:16 |
|---|---|---|---|---|
| serum(n=5) | 89-96% | 79-86% | 94-106% | 92-101% |
| EDTA plasma(n=5) | 88-100% | 89-98% | 81-87% | 81-87% |
| Heparin plasma(n=5) | 88-96% | 95-106% | 96-103% | 92-102% |