| 48 T | $265.00 |
| 96 T | $376.00 |
| 96 T*5 | $1598.00 |
| Concentration (ng/mL) | OD | Corrected OD |
|---|---|---|
| 500.00 | 2.738 | 2.660 |
| 250.00 | 1.742 | 1.664 |
| 125.00 | 1.014 | 0.936 |
| 62.50 | 0.764 | 0.686 |
| 31.25 | 0.400 | 0.322 |
| 15.63 | 0.280 | 0.202 |
| 7.82 | 0.123 | 0.045 |
| 0.00 | 0.078 | 0.000 |

Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
| Matrix | Recovery range | Average |
|---|---|---|
| serum(n=5) | 91-102% | 97% |
| EDTA plasma(n=5) | 81-95% | 88% |
| Heparin plasma(n=5) | 80-89% | 85% |
| Matrix | 1:2 | 1:4 | 1:8 | 1:16 |
|---|---|---|---|---|
| serum(n=5) | 79-92% | 79-85% | 80-90% | 91-97% |
| EDTA plasma(n=5) | 88-96% | 87-93% | 88-95% | 88-95% |
| Heparin plasma(n=5) | 94-108% | 95-103% | 83-90% | 87-94% |