| 48 T | $320.00 |
| 96 T | $458.00 |
| 96 T*5 | $1947.00 |
| Concentration (ng/mL) | OD | Corrected OD |
|---|---|---|
| 10.00 | 3.137 | 3.077 |
| 5.00 | 2.153 | 2.093 |
| 2.50 | 1.275 | 1.215 |
| 1.25 | 0.720 | 0.660 |
| 0.63 | 0.479 | 0.419 |
| 0.32 | 0.276 | 0.216 |
| 0.16 | 0.128 | 0.068 |
| 0.00 | 0.060 | 0.000 |

Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
| Matrix | Recovery range | Average |
|---|---|---|
| serum(n=5) | 92-104% | 98% |
| EDTA plasma(n=5) | 82-92% | 87% |
| Heparin plasma(n=5) | 83-90% | 87% |
| Matrix | 1:2 | 1:4 | 1:8 | 1:16 |
|---|---|---|---|---|
| serum(n=5) | 80-92% | 82-92% | 88-96% | 93-101% |
| EDTA plasma(n=5) | 77-86% | 90-98% | 95-104% | 95-104% |
| Heparin plasma(n=5) | 83-96% | 93-105% | 90-103% | 96-104% |