| 48 T | $265.00 |
| 96 T | $376.00 |
| 96 T*5 | $1598.00 |
| Concentration (ng/mL) | OD | Corrected OD |
|---|---|---|
| 10.00 | 2.694 | 2.639 |
| 5.00 | 1.746 | 1.691 |
| 2.50 | 1.264 | 1.209 |
| 1.25 | 0.736 | 0.681 |
| 0.63 | 0.460 | 0.405 |
| 0.32 | 0.231 | 0.176 |
| 0.16 | 0.154 | 0.099 |
| 0.00 | 0.055 | 0.000 |

Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
| Matrix | Recovery range | Average |
|---|---|---|
| serum(n=5) | 80-94% | 87% |
| EDTA plasma(n=5) | 85-91% | 88% |
| Heparin plasma(n=5) | 92-105% | 99% |
| Matrix | 1:2 | 1:4 | 1:8 | 1:16 |
|---|---|---|---|---|
| serum(n=5) | 96-105% | 87-101% | 93-102% | 88-96% |
| EDTA plasma(n=5) | 78-90% | 78-84% | 83-96% | 83-96% |
| Heparin plasma(n=5) | 86-96% | 89-100% | 78-85% | 90-101% |