| 48 T | $320.00 |
| 96 T | $458.00 |
| 96 T*5 | $1947.00 |
| Concentration (pg/mL) | OD | Corrected OD |
|---|---|---|
| 2000.00 | 2.613 | 2.561 |
| 1000.00 | 1.719 | 1.667 |
| 500.00 | 0.896 | 0.844 |
| 250.00 | 0.536 | 0.484 |
| 125.00 | 0.283 | 0.231 |
| 62.50 | 0.144 | 0.092 |
| 31.25 | 0.115 | 0.063 |
| 0.00 | 0.052 | 0.000 |

Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
| Matrix | Recovery range | Average |
|---|---|---|
| serum(n=5) | 81-87% | 84% |
| EDTA plasma(n=5) | 92-99% | 96% |
| Heparin plasma(n=5) | 82-94% | 88% |
| Matrix | 1:2 | 1:4 | 1:8 | 1:16 |
|---|---|---|---|---|
| serum(n=5) | 88-102% | 88-101% | 93-107% | 96-102% |
| EDTA plasma(n=5) | 95-105% | 80-93% | 83-92% | 83-92% |
| Heparin plasma(n=5) | 93-105% | 96-102% | 77-86% | 92-101% |