| 48 T | $320.00 |
| 96 T | $458.00 |
| 96 T*5 | $1947.00 |
| Concentration (ng/mL) | OD | Corrected OD |
|---|---|---|
| 10.00 | 2.579 | 2.524 |
| 5.00 | 1.696 | 1.641 |
| 2.50 | 0.959 | 0.904 |
| 1.25 | 0.523 | 0.468 |
| 0.63 | 0.266 | 0.211 |
| 0.32 | 0.130 | 0.075 |
| 0.16 | 0.093 | 0.038 |
| 0.00 | 0.055 | 0.000 |

Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
| Matrix | Recovery range | Average |
|---|---|---|
| serum(n=5) | 78-84% | 81% |
| EDTA plasma(n=5) | 85-91% | 88% |
| Heparin plasma(n=5) | 92-105% | 99% |
| Matrix | 1:2 | 1:4 | 1:8 | 1:16 |
|---|---|---|---|---|
| serum(n=5) | 96-107% | 93-102% | 93-99% | 89-96% |
| EDTA plasma(n=5) | 86-96% | 79-88% | 83-95% | 83-95% |
| Heparin plasma(n=5) | 87-94% | 94-107% | 80-94% | 88-95% |